Abstract 1227P
Background
Assessment of HER2 status is critical for diagnosis and guiding therapy in breast cancer. However, the current immunohistochemistry-based analysis of HER2 requires tissue biopsy, is time-consuming, subjective, and semi-quantitative. This is the reason for delayed or incorrect treatment decisions and worse outcomes. To address these limitations, we developed an innovative inPROBE diagnostic device, which allows for real-time quantitative and operator-independent HER2 level determination, directly in the patient’s body.
Methods
The HER2 measurements were performed using the inPROBE device consisting of an optical sensor designed for the detection of the soluble HER2 or the anti-HER2 drug – Trastuzumab (TMAB). The in vitro inProbe testing was done on the standard HER2 protein solutions to demonstrate the device working range and the tissue accumulation of TMAB was assessed in vivo on the mouse xenograft model of a human HER2+ cancer. The device safety was confirmed on rats with a probe insertion into the mammary glands to monitor the clinical, biochemical and pathological changes.
Results
inProbe working range tested on HER2 solution was within the concentration of pg/ml – mg/ml with good linearity and selectivity toward HER2. In preclinical safety experiments, inPROBE insertion into the mammary glands did not cause any adverse clinical signs, changes in blood parameters, or histology. In alternative sets of experiments, the TMAB-inProbe proved its utility for direct detection of the anti-HER2 drug accumulating in the HER2+ tumor tissue.
Conclusions
HER2-inPROBE is a specific and sensitive sensor, allowing for rapid, quantitative analysis of HER2. Given its high adaptability, the inPROBE platform presents the potential to be used for the detection of various diseases biomarkers, without the need for tissue biopsy. The clinical trial to determine the diagnostic accuracy and safety of the inPROBE in 192 patients with early breast cancer is underway (NCT05415943).
Clinical trial identification
NCT05415943.
Editorial acknowledgement
Legal entity responsible for the study
SDS Optic.
Funding
STRATEGMED II/269364/5/NCBR/2015; EU, Horizon 2020 SME Instrument (No. 783818).
Disclosure
M. Staniszewska, M. Śmietana, M. Staniszewski: Financial interests, Institutional, Membership or affiliation: SDS Optic, Inc. All other authors have declared no conflicts of interest.
Resources from the same session
1219P - Artificial intelligence-based breast cancer detection facilitates automated prognosis marker assessment using multiplex fluorescence immunohistochemistry
Presenter: Tim Mandelkow
Session: Poster session 14
1220P - Comprehensive diagnose of programmed death-ligand 1 from two-dimensional to three-dimensional in breast cancer with computer-aided artificial intelligence system
Presenter: Yi-Hsuan Lee
Session: Poster session 14
1221P - The functional domain of BRCA1/2 pathogenic variants (PVs) as potential biomarkers of second tumor and domain-related sensitivity to PARP-inhibitors
Presenter: Lorena Incorvaia
Session: Poster session 14
1222P - Detection of androgen-receptor splice variant 7 messenger RNA in circulating tumor cells of prostate cancer by in vitro assay
Presenter: Hoin Kang
Session: Poster session 14
1223P - Homologous recombination deficiency (HRD) testing on ovarian cancer ascites: A feasibility study
Presenter: Alberto Ranghiero
Session: Poster session 14
1224P - Detection of circulating tumor DNA (ctDNA) in untreated patients (pts) with cancer: Implications for early cancer detection (ECD)
Presenter: Yoshiaki Nakamura
Session: Poster session 14
1225P - Combining ctDNA and tissue-based-genomic profiling in advanced cancer: A real-world evidence prospective study in non-Western patients treated at Gustave Roussy cancer campus
Presenter: Tony Ibrahim
Session: Poster session 14
1226P - Multi-site validation of a deep learning solution for HER2 profiling of breast cancer from H&E-stained pathology slides
Presenter: Salim Arslan
Session: Poster session 14
1228P - A circulating tumor cell (CTC) based assay for diagnostic immunocytochemistry profiling of lung cancer
Presenter: Nitesh Rohatgi
Session: Poster session 14
1229P - Selective phenotypic and genotypic evaluation of circulating glial cells for improved diagnosis of glial malignancies
Presenter: Sewanti Limaye
Session: Poster session 14