Oops, you're using an old version of your browser so some of the features on this page may not be displaying properly.

MINIMAL Requirements: Google Chrome 24+Mozilla Firefox 20+Internet Explorer 11Opera 15–18Apple Safari 7SeaMonkey 2.15-2.23

Poster viewing and lunch

39P - A validation study of a self-testing capillary kit, the rhelise™ kit for therapeutic dose monitoring (TDM) of tamoxifen, Z-endoxifen, and 4-hydroxytamoxifen in breast cancer patients

Date

12 May 2023

Session

Poster viewing and lunch

Presenters

Elham Hedayati

Citation

Annals of Oncology (2023) 8 (1suppl_4): 101218-101218. 10.1016/esmoop/esmoop101218

Authors

E. Hedayati1, I. Shabo2, P. Rydberg3, S. Rehnmark4, H. Randahl4, A. Lindqvist4, P.F. Hall5, J. Bergqvist3, Y. Wengstrom6

Author affiliations

  • 1 Karolinska Institutet - Bioclinicum, Solna/SE
  • 2 Karolinska University Hospital-Solna, Solna/SE
  • 3 Karolinska Institute, Stockholm/SE
  • 4 Redhot Diagnostics AB, Sodertalje/SE
  • 5 Karolinska University Hospital and Karolinska Institute - Huddinge, Stockholm/SE
  • 6 Karolinska Institutet, Huddinge/SE

Resources

Login to get immediate access to this content.

If you do not have an ESMO account, please create one for free.

Abstract 39P

Background

Tamoxifen is an effective drug reducing the risk of dying from breast cancer. This study aims to validate and study the feasibility of serial assessments, including therapeutic drug monitoring of tamoxifen (TAM), 4-hydroxytamoxifen (4HT) and Z-endoxifen (Z-END) by capillary blood sampling.

Methods

A prospective, single-centre study including women with stage 0-3 breast cancer receiving adjuvant tamoxifen 20 mg/day for a minimum of 2 months. Study design: Blood samples for repeated measurement of TAM, 4HT and Z-END were drawn capillary in total at 4-time points, at inclusion (baseline), and after 1, 2, and 3 weeks for each participant; and venously in total at two time-points, at inclusion (baseline), and after 3 weeks for each participant. At baseline, the capillary blood test concentrations of TAM, 4HT and Z-END were taken first by the research nurse and second by the patient. At each time, participants will be asked to leave 1 vial of capillary blood using the rhelise™ kit and 2 samples of conventional venous blood for blood and plasma. Primary Objective To validate the rhelise™ kit for monitoring TAM, 4HT and Z-END among patients who have ongoing adjuvant tamoxifen by testing for equivalence between concentrations found in the capillary sample and the venous blood sample (gold standard). Primary endpoint Blood concentrations and correlations of TAM, 4HT and Z-END at baseline and 3 weeks by capillary and venous blood sampling (whole blood/plasma). Secondary objectives To validate user acceptability and feasibility of self-testing the capillary kit by comparing the capillary blood test concentrations of TAM, 4HT and Z-END taken by the patient to the capillary sample taken by the research nurse. Secondary endpoints Capillary blood test concentrations of TAM, 4HT and Z-END were taken by the patient and the research nurse at baseline.

Results

A capillary sampling kit was used for 40 participants. Mean TAM, 4HT and Z-END concentrations did not differ significantly in the 3 types of samples. Mean capillary TAM, 4HT and Z-END concentrations did not differ significantly between nurse and patient.

Conclusions

TDM of TAM using capillary blood sampling is feasible.

Clinical trial identification

NCT05133674.

Legal entity responsible for the study

Elham Hedayati, Karolinska University Hospital.

Funding

Vinnova & Karolinska University Hospital, Karolinska Institutet.

Disclosure

S. Rehnmark, H. Randahl: Financial Interests, Personal and Institutional, Ownership Interest: Redhotdiagnostics AB. A. Lindqvist: Financial Interests, Personal and Institutional, Full or part-time Employment: Redhotdiagnostics AB. All other authors have declared no conflicts of interest.

This site uses cookies. Some of these cookies are essential, while others help us improve your experience by providing insights into how the site is being used.

For more detailed information on the cookies we use, please check our Privacy Policy.

Customise settings
  • Necessary cookies enable core functionality. The website cannot function properly without these cookies, and you can only disable them by changing your browser preferences.